BW Technologies MicroClip XT Manual de usuario Pagina 42

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 57
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 41
GasAlertMicroClip
User Manual
Verification
Peak exposure level (ppm or %);
After calibration is complete and the detector is in normal
operating mode, test it using a gas cylinder other than the
one used in calibration. The gas concentration should not
exceed the sensor's detection range. Confirm that the LCD
shows the expected concentration. To ensure the readings
are accurate, apply the test gas for the same amount of time
as was applied to the sensor when it was calibrated.
Alarm duration in seconds; and
Alarm type.
Downloading Datalogs and Event Logs
The datalog and event log files can only be downloaded to a
PC using an IR device (e.g., IR Link Adapter or MicroDock II
Base Station).
Datalogs
Maintenance
The datalogger allows the detector to record various
information so you can compile a report. The detector has a
sample rate of 15-second intervals. The detector is capable
of storing 16 hours of information. When the memory is full,
the detector replaces the oldest data with the most recent
data.
To keep the detector in good operating condition, perform
the following basic maintenance as required:
Calibrate, bump check, and inspect the detector at
regular intervals.
Event Logs
Keep an operations log of all maintenance, bump checks,
calibrations, and alarm events.
The event logging feature keeps a record of the ten most
recent gas alarm events. The information that is recorded
from an event is as follows:
Clean the exterior with a soft damp cloth. Do not use
solvents, soaps, or polishes.
Serial number;
Gas type;
Start time of event;
34
Vista de pagina 41
1 2 ... 37 38 39 40 41 42 43 44 45 46 47 ... 56 57

Comentarios a estos manuales

Sin comentarios