GasAlertMicroClip
Operator’s Manual
14
Alarm Display Alarm Display
Sensor Alarm
• During startup Error [sensor name]
displays
• During normal operation Err displays
Confidence Beep and IntelliFlash
• One beep and flash every second
Note: Confidence Beep and
IntelliFlash automatically
deactivate during low battery
alarm, self-test fail, calibration fail,
bump test fail, or an alarm event.
Low Battery Alarm
• Sequence of 10 rapid sirens and
alternating flashes with 7 seconds of
silence in between (continues for
15 minutes)
• and
L
flash,
LOW BAT
displays, and the vibrator alarm activates
• After 15 minutes of the Low Battery
alarm, the Automatic Shutdown Alarm
sequence begins
• OFF displays before deactivating
Automatic Shutdown Alarm
• Sequence of 10 rapid sirens and
alternating flashes with 1 second of
silence in between (sequence reactivates
seven times)
• LOW BAT and L display, and
vibrator alarm activates
• OFF displays before deactivating
Note: Alarms can be set to latching or non-latching. To enable/disable
latching, select/deselect Latching Alarms in Fleet Manager II. Local
regulations may require a latching alarm.
If Low Alarm Acknowledge is enabled and a low alarm occurs, press
C to disable the audible alarm. The visual and vibrator alarms remain
activated. If the alarm escalates to a high, STEL, or TWA alarm, the
audible alarm reactivates.