Safety Information - Read First
aCautions
⇒ Warning: Substitution of components may impair Intrinsic Safety.
⇒ Caution: For safety reasons, this equipment must be operated and serviced by qualified personnel only. Read
and understand the user manual completely before operating or servicing.
⇒ Do not use the detector if it is damaged. Before you use the detector, inspect the case. Look for cracks or
missing parts.
⇒ If the detector is damaged or something is missing, contact
BW Technologies by Honeywell immediately.
⇒ Use only sensor(s) specifically designed for your GasAlertMicroClip model. (See the section,
Replacement Parts
and Accessories).
⇒ Calibrate the detector before first-time use and then on a regular schedule, depending on use and sensor
exposure to poisons and contaminants. BW recommends at least once every 180 days (6 months).
⇒ BW recommends to “bump test” the sensors, before each day’s use, to confirm their ability to respond to gas by
exposing the detector to a gas concentration that exceeds the alarm setpoints. Manually verify that the audible
and visual alarms are activated. Calibrate if the readings are not within the specified limits.
⇒ It is recommended that the combustible sensor be checked with a known concentration of calibration gas after
any known exposure to catalyst contaminants/poisons (sulfur compounds, silicon vapors, halogenated
compounds, etc).
⇒ The combustible sensor is factory calibrated to 50% LEL methane. If monitoring a different combustible gas in
the % LEL range, calibrate the sensor using the appropriate gas.
⇒ Caution: High off-scale readings may indicate an explosive concentration.
⇒ Only the combustible gas detection portion of this instrument has been assessed for performance by CSA
International.