GasAlertMicroClip
Troubleshooting
Troubleshooting
If you are still unable to correct the problem, contact
BW Technologies by Honeywell.
The detector’s electronics are protected from variations in
humidity and corrosive atmospheres. If you encounter a
problem, try the solutions listed in the following table.
Table 12. Troubleshooting Tips
Problem Possible Cause Solution
The detector does not activate.
→ Depleted battery
→ Damaged or defective detector
→ Charge battery (see
Charging the
Detector)
→ Contact
BW Technologies by
Honeywell
The detector enters alarm mode
immediately when activated.
→ Sensor needs to stabilize
→ Low battery alarm
→ Detector requires calibration
→ Used sensor: wait 60 seconds
New sensor: wait 5 minutes
→ Charge battery (see
Charging the
Detector)
→ Calibrate detector (see
Calibration)
The activation self-test fails.
→ General fault
→ Sensor failure
→ Contact
BW Technologies by
Honeywell
→ Replace sensor (see
Replacing a
Sensor or Sensor Filter)
39