GasAlertMicroClip XT
Technical Reference Guide
16
Self-Test Pass
When the detector has passed all startup self-tests, it enters normal
operation. The LCD displays the ambient gas readings.
The detector automatically begins
recording the peak (maximum) gas exposure,
calculating the short-term exposure level (STEL), and
calculating the time-weighted average (TWA) exposures.
Self-Test Fail
If the following error message displays after entering normal operation,
refer to Startup Troubleshooting
.
Battery Test
The battery is tested when the detector is activated and continuously
thereafter. A newly charged battery will typically operate for 10-12
hours.
Battery power is continually displayed during normal operation. If
battery power is low, L and flashes. The detector per-
forms a sequence of 10 rapid sirens and alternating flashes with 7
seconds of silence in between (continues for 15 minutes).
If battery power becomes critically low, L and
LOW BAT dis-
play. The detector performs a sequence of 10 rapid sirens with 1
second of silence in between (sequence reactivates seven times).
The detector then displays
OFF and the detector deactivates.
Note
If enabled, Confidence Beep and IntelliFlash automatically
deactivate during a low battery alarm. Refer to Confidence
Beep and IntelliFlash.
Backlight
The backlight automatically activates
during startup,
when the pushbutton is pressed (then deactivates after 10 sec-
onds), and
when there is an alarm condition (unless Stealth is enabled).