55
GasAlertMicroClip XT
Troubleshooting
Table 14. Troubleshooting
Problem Possible Cause Solution
Detector intermittently enters
alarm without reason.
Ambient gas levels are near alarm setpoint or
the sensor is exposed to a puff of the target
gas
Detector is operating normally. Use caution in
suspect areas. Check the peak (maximum) gas
exposure reading.
Alarm setpoints defined incorrectly Reset alarm setpoints. Refer to Factory Gas
Alarm Setpoints and Sensor Configuration.
Detector requires calibration Calibrate the sensors. Refer to Calibration.
Missing or faulty sensor(s) Replace the sensor. Refer to Replacing a Sensor
or Sensor Filter.
Features and options are not
operating as expected.
Changes have been made in Fleet Manager II Verify settings in Fleet Manager II are correct.
Charging
Battery has been charging for
3+ hours. The charging indicator
on the detector LCD shows the
battery is still charging.
Battery is trickle charging Battery is fully charged and ready for operation.
Battery indicator does not display
when charging.
Detector is depleted below normal levels Charge the battery for 8 hours. Detector LEDs
may light during first 5 hours. This is normal. If
the battery indicator does not light after charging
for 8 hours, contact BW Technologies by Honey-
well.
When detector is activated after
charging, the battery indicator does
not display.
Battery is defective Contact BW Technologies by Honeywell
.