Calibration Troubleshooting
Calibration Troubleshooting
Error Screen Problem Solution Error Screen Problem Solution
Auto-zero
Unsuccessful
H
2
S, CO, or LEL sen-
sor fails to auto-zero,
or O
2
sensor fails to
calibrate.
Attempt calibration again.
Refer to Calibration
. If
ERROR displays again,
replace the sensor. Refer to
Replacing a Sensor or Sen-
sor Filter.
No Gas Detected
If the applicable gas is
not detected within 2
minutes, the detector
fails the sensor.
Ensure the sensor is enabled.
Verify gas cylinder is not
empty or past the expiration
date. Check/replace the regu-
lator.
Attempt calibration again. If
the sensor fails the span a
second time, refer to Replac-
ing a Sensor or Sensor Filter.
Auto Span
Unsuccessful
H
2
S, CO, or LEL sen-
sor fails to auto-zero,
or O
2
sensor fails to
calibrate.
Ensure sensor is enabled.
Verify gas cylinder is not
empty or past the expiration
date. Check/replace the
regulator. Attempt calibra-
tion again. If the sensor fails
the span again, refer to
Replacing a Sensor or Sen-
sor Filter.
Calibration Due Date
Overdue
A sensor displays a
negative number for a
next due date after
calibration is per-
formed.
Calibration for the sensor was
unsuccessful. The due date
will not reset. Attempt calibra-
tion of the sensor again. If still
unsuccessful, refer to Replac-
ing a Sensor or Sensor Filter.
Calibrate the new sensor
immediately.
IR Lock Enabled
IR-Lock displays
when calibration is
attempted.
Perform calibration using
the IR Link with Fleet Man-
ager II software, or insert
the detector into the
MicroDock II station. Refer
to Cal Lock (Calibration IR
Lock) in User Options and
Calibration
.