
45
GasAlertMicroClip XT
Calibration
Verification
1. After calibration is complete and the detector returns to
normal operation, verify the calibration using a gas cylin-
der other than the one used for calibration.
2. The gas concentration should not exceed the sensor's
detection range. Confirm the LCD shows the expected
concentration.
3. To ensure the readings are accurate, apply the verifica-
tion gas for the same amount of time as was applied to
the sensor when it was calibrated.
Example: H
2
S span time 2 minutes therefore, apply verification
gas for 2 minutes.
Comentarios a estos manuales