GasAlertMicroClip XT
Technical Reference Guide
52
5. Carefully replace the circular rigidified flex PCB atop the
metal sensor posts. Ensure the purple plastic sensor post
is inserted into the clear plastic hole. Take care not to
tear the flex cable.
6. Press down to secure the circular rigidified flex PCB atop
the metal sensor posts.Take care not to press down too
hard and accidentally activate the detector.
7. To complete the detector, refer to Reassembling the
detector.
Reassembling the detector
1. To re-assemble the detector, perform the following:
Verify the PCB is seated correctly and inserted exactly as it
was removed (sensors facing the front shell).
Replace the two PCB screws.
Visually inspect the battery to ensure no damage has occurred.
When replacing the rear shell, ensure the charging pins (bot-
tom of inside rear shell) are aligned with the corresponding
holes on the PCB. If the contact pins are bent, the battery will
not charge correctly.
Ensure the rib on the interior rear shell inserts between the bat-
tery and the PCB.
Figure 13. Replacing the Rear Shell
Press the front and rear shells together firmly to ensure a
proper seal. Ensure the front and rear shell have a uniform,
tight 1/16 in (1 mm) seal on all sides of the detector.
When replacing the screws, they must be seated properly to
prevent cross threading. Turn the screw counter-clockwise until
a click is heard and then begin tightening the screw clockwise.
2. New sensors must be calibrated. Activate the detector
and calibrate the sensor(s). Refer to Calibration
.