GasAlertMicroClip XT
Technical Reference Guide
Startup Troubleshooting
Error Screen Problem Solution Error Screen Problem Solution
Sensor Error
The sensor failed dur-
ing the self-test.
Calibrate the sensor(s).
Refer to Calibration
. Reacti-
vate the detector. If error
displays again, replace the
sensor. Refer to Replacing
a Sensor or Sensor Filter.
IR Lock Enabled
If the IR Lock screen
displays, an IR device
is required to calibrate
the sensors.
Perform calibration using the
IR Link with Fleet Manager II
software, or insert the detector
into the MicroDock II station.
Refer to Cal Lock (Calibra-
tion IR Lock) in User Options
and Calibration
.
Calibration
Overdue
Displays when cali-
bration is overdue. If
the Force Calibra-
tion When Overdue
option is enabled, the
sensor(s) must be
calibrated to enter
normal operation.
Press C to continue and
calibrate the sensor(s)
immediately. Refer to
Calibration
.
If the IR Lock enabled
screen displays, the
MicroDock II station or the
IR Link with Fleet Manager
II must be used to calibrate.
Bump Test Fail
A bump test has just
been performed. The
detector is prompting
for another bump test
because a sensor(s)
has failed.
Perform another Bump Test
.
Ensure the cylinder is not
empty and that the cylinder is
not past the expiry date.
Ensure the regulator is fully
opened to apply gas. If Bump
Check Today displays again,
calibrate the sensors. Refer to
Calibration
. If the calibration is
unsuccessful, refer to Replac-
ing a Sensor or Sensor Filter.
Forced Calibration
If Force Calibration
When Overdue is
enabled, the sensors
must be calibrated to
enter normal opera-
tion.
Press and hold C to cali-
brate the sensors, or press
C and release to deactivate
the detector. Refer to Cali-
bration. If the IR Lock
enabled screen displays,
an IR device must be used
to calibrate.
Sensor Fail
A sensor has failed
during the startup self-
test.
Perform a Bump Test
and
reactivate the detector. If the
sensor fails again, perform
Calibration
. Reactivate the
detector again. If the sensor
still does not pass, refer to
Replacing a Sensor or Sensor
Filter.