GasAlertMicroClip XT
Calibration
8. Ensure the sensors to be calibrated are enabled in Fleet
Manager II.
9. Using , select the concentration value(s) in the
Cal Gas field for each sensor. The values entered in
Fleet Manager II must match the gas concentration val-
ues on the gas cylinder.
10. Click to save the settings to the
detector.
Calibrating with the IR Link
a Caution
Only calibrate in a fresh air environment and in a safe area. Do
not calibrate the detector during or immediately after
charging.
EC To calibrate the detector with the IR Link, complete the following pro-
cedure:
1. Complete steps #1-10 under Setting Span Gas Concen-
tration Values.
2. Click on the Device Operations tab.
3. Click
Calibrate. The detector begins calibration. Refer to
Auto Zero and Oxygen Sensor Calibration
.
Calibration Procedure
a Caution
Only calibrate in a fresh air environment and in a safe area. Do
not calibrate the detector during or immediately after
charging.
1. Press and hold C. The detector performs the OFF count-
down. Continue holding C as the detector briefly deacti-
vates.
2. The detector then reactivates and performs the CAL
countdown. Continue holding C until the CAL count-
down is complete.
Note
If C is not held for the entire countdown, the detector will deac-
tivate.