GasAlertMicroClip XT
Maintenance
Replacing a Sensor or Sensor Filter
a Warning
To avoid personal injury, only use sensors that are
specifically designed for the detector. Refer to Replacement
Parts and Accessories.
Use proper ESD handling practices.
Each sensor has a high degree of resistance to common vapors
and gases. To clear a sensor, move the detector to a non-hazard-
ous environment and wait 10 to 30 minutes.
Do not expose a sensor to vapors of inorganic solvents such as
fumes from paint thinners, or organic solvents such as benzoic
acids and acrylic acids.
Ensure hands are clean or wear gloves before handling compo-
nents.
To replace a sensor or sensor filter, refer to
Table 13.
,
Figure 11.
,
Figure 13.
, and
the following procedures.
Figure 11. Replacing a Sensor or Sensor Filter