BW Technologies MicroClip XT Especificaciones Pagina 53

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 74
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 52
GasAlertMicroClip XT
Technical Reference Guide
42
Auto Zero and Oxygen Sensor Calibration
Note
Do not apply calibration gas until APPLY GAS displays, other-
wise the auto zero function will fail.
3. flashes while the detector zeroes the combusti-
ble and toxic sensors, and calibrates the oxygen sensor.
When auto zero is complete, the detector beeps twice.
Auto Zero Successful: If the sensors successfully zero, the detector
proceeds to the Auto Span
function.
Auto Zero Unsuccessful: If the sensors fail auto zero, an error mes-
sage displays showing which sensor failed. Refer to Calibration Trou-
bleshooting.
Auto Span
4. When auto zero is complete, APPLY GAS displays.
5. Attach the calibration cap to the detector. Refer to Figure
8. Open the valve on the regulator and apply gas at a
flow rate of 250-500 ml/min.
K flashes and displays.
Vista de pagina 52
1 2 ... 48 49 50 51 52 53 54 55 56 57 58 ... 73 74

Comentarios a estos manuales

Sin comentarios